How to Calibrate the T36 Colortrac Scanner by Global Scanning

Article ID: ART173772 | Date published: 02/28/2019 | Date last updated: 02/28/2019
 

Description

How to Calibrate the T36 Colortrac Scanner by Global Scanning.

Solution

How to Calibrate the T36 scanner

Ensure the scanner is cleaned before the Calibration is performed. Calibrating a dirty scanner can introduce more problems. Remove the Document Return Guides.

From the Home screen, select the scanner "i" in the bottom left hand corner.

Press Calibrate.

Insert the calibration target into the scanner black end first face up. Press OK.

The calibration target will feed back & forth.


 

Wait a few minutes until the process is complete. When complete the software will automatically return to the Home page.  Press rewind on the scanner to eject the target. Replace the calibration target in its packaging and keep in a safe place where it will not be damaged or bent until required again.

Auto-Stitch the scanner

Ensure the scanner is cleaned and calibrated before the Auto-stitch is performed.  Remove the Document Return Guides.

From the Home screen, select the scanner "i" in the bottom left hand corner.

Press Auto Stitch.

Insert the calibration target into the scanner black end first face up. Press OK.

The calibration target will feed back & forth.

 

Wait a few minutes until the process is complete. When complete the software will automatically return to the Home page.

Press rewind on the scanner to eject the target. Replace the calibration target in its packaging and keep in a safe place where it will not be damaged or bent until required again.